There are many photosensitizers that are presently in their advanced
stages of clinical development. All compounds being studied share characteristics
that are required of an ideal photosensitizer.
Tissue
specificity of a photosensitizer is vital to PDT. Compounds have been
selected based upon their ability to be targeted to specific tissue
or ability to be loosely bonded to unreactive compounds that can be
targeted. Non toxicity of the compound is also essential. Due to restrictions
in accessibility of today’s light delivering systems and the depth
that this light can penetrate tissue has provided yet another indirect
requirement of photosensitizers. Wavelengths which diffuse through the
tissue most effectively are 640 nm and 700-800 nm and can penetrate
to a depth of 2-3 mm and 5-6 mm respectively. Therefore the compounds
of interest must be able to absorb wavelengths within this range of
the visible spectrum. Upon being activated the photosensitizer should
not dissociate or cause any undesired side effects.
The
efficiency of a photosensitizer is measured via molar absorption coefficient
(e). The molar absorption coefficient “(formerly known as the
molar extinction coefficient), is a property of a molecule undergoing
an electronic transition” (3). It is a fixed constant and is unique
to every molecule. The Beer-Lambert law; A= e c l, relates molar absorptivity
e to Absorbance A (or optical density), molar concentration c and length
l of the sample cell (with respect to a cuvette length in a lab). Molecules
with a higher molar absorption coefficient are desirable for PDT.
All
compounds have the ability to absorb light, but what makes photosensitizers
unique is the method by which they decay to their ground state. Energy
E of a specific wavelength can be described by E = h c / lamda, where
h is Plank’s constant, c is the speed of light and lamda is a
single wavelength. The relationship that wavelength shares with energy
allow molecules to become electronically excited when exposed to light
(at specific wavelengths). Once in this excited state excess energy
can be expressed through two possible transitions, spin-allowed transitions
or via spin-prohibited transitions. (2) Spin allowed transitions encompass
internal conversion; A method of lowering energy by means conversion
into heat (kinetic energy) through molecular vibration, rotations and
collision with other local molecules.(1) Spin allowed transitions also
include fluorescence which follow vibrational relaxation of the excited
molecule. The second type of transition, spin-prohibited transition
is of particular interest and is fundamental to the concept of PDT.
Spin prohibited transitions involve relaxation of an excited molecule
via phosphorescence and more importantly by intersystem crossing. Intersystem
crossing is method by which an activated molecule reaches a new electronic
state (triplet state) which is highly reactive and can trigger photochemical
reactions. What makes photosensitizers unique is that they have a higher
probability of reaching this reactive triplet state.